SAFETY PRECAUTIONS

Electrical Safety

The USB port output connector must ONLY be connected to a battery powered notebook computer with no other wired connections to any other device. The laptop computer must only be ever powered form a mains supply using a IEC 60601-1 certified power supply when used with a patient

DO NOT use near water, wet locations or outdoors.

DO NOT OPERATE IN THE PRESENCE OF FLAMMABLE MATERIALS. POSSIBLE EXPLOSION HAZARD EXISTS. This is particularly relevant to uses in operating theatres where inflammable anaesthetic gases may be present.

Certification

This product meets the provisions of Council Directive 93/42/EEC

(Medical Devices Directive) Annex VII. It complies with standards:

EN 60601-1:2006 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance