Guidance and manufacturer’s declaration – electromagnetic emissions

Guidance and manufacturer’s declaration – electromagnetic emissions

The CapiScope HVCS is intended for use in the electromagnetic environment specified below. The customer or the user of the CapiScope HVCS should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environmentguidance

RF emissions

CISPR 11

Group 1

The CapiScope HVCS uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions

CISPR 11

Class A

The CapiScope HVCS is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Not applicable

Voltage fluctuations/ flicker emissions

IEC 61000-3-3

Not applicable

Guidance and manufacturer’s declaration – electromagnetic immunity

The CapiScope HVCS was not tested for IMMUNITY to ELECTROMAGNETIC DISTURBANCES.